March 20-March 26 2018
This week the Power Puff Girls worked to accomplish testing and redesign of the recirculation apparatus. The testing was done for three rounds using 5 concentrations of approximately .17g/L. the results showed an average yield of 65.11%. This value accounts for the mass of nanoparticles that collected on the magnet and were removed from the system. Along with this, we had to consider the control. The control test was run my flowing the same volume through the system without any magnetic field. After the total 5L was run through the system, the number of nanoparticles left in the tubing and bags was then flushed out of the system and weighed. The percent of nanoparticles left in the system was 31.39%. This percentage is large and is mostly due to the nanoparticles attaching to the walls of the bag and tubing. The percent needs to be accounted for when considering the results from testing if the amount left in the system is approximately 30%, then we would assume that the number of nanop...